Price of the test:
Please enquire our customer service at sample.receipt@caiqtest.co.nz
Acceptable Matrices:
Our accreditation covers: milk, milk powders, infant formula
Limit of Detection:
Any concentration greater than 0.008 % is detectable.
Sample Size Requirements:
50 grams of sample is needed.
Equipment:
Automatic Titrator
Reportable Units:
%
Turn Around Time:
7 Business Days
Is Urgent Service Available?
No (Enquire our customer service for special requirements at sample.receipt@caiqtest.co.nz)
Method Reference:
GB 5009.44-2016 method I & III
Method Description:
Testing reports values for total chloride. A portion of homogenized sample is mixed with potassium ferrocyanide solution and zinc acetate solutions for protein precipitation. The supernatant of extract is analyzed on automatic titrator. The amount of titrant (AgNO3) consumed in the analysis can be used to calculate the amount of total chloride in the sample.